YH2000 Thermocouple and Thermal Resistance Temperature Automatic Calibration Device
YH2000 Thermocouple and Thermal Resistance Temerature Automatic Calibration Device
YH2000 temperature (thermocouple, thermal resistance) automatic calibration device is designed and manufactured strictly in accordance with relevant national calibration regulations. It is used to calibrate various working thermocouples and thermal resistance, first-class and second-class standard thermocouples, temperature secondary instruments, expansion thermometers and other temperature instruments, as well as various temperature sensors and temperature transmitters. Fully automated operation, automatic temperature regulation, measurement channel control, data acquisition and processing, report and certificate output, and other functions.
◆In line with the latest national verification regulations and norms;
◆Thermocouple type: S, R, B, K, N, E, J, T, EA-2, etc.;
◆Thermal resistance type: Pt 1000, Pt 100, Pt 10, Cu 100, Cu 50, BA1, BA2, G (two-wire, three-wire, four-line);
◆Type of expansion thermometer: glass liquid thermometer, bimetal thermometer, pressure thermometer;
◆Special low potential wiring terminal, parasitic thermal potential long-term guarantee not to exceed ± 0.2 μV;
◆The recording curve includes the temperature, power and power change curve;
◆Thermocouple and thermal resistance can be verified at the same time, and can be verified in the oil bath;
◆Compatible with thermocouple and thermal resistance wiring, four-wire resistance measurement method completely eliminate the external lead resistance;
◆Reliable temperature control, intelligent P I D regulation system, good constant temperature stability;
◆It can control the verification furnace and oil (water) bath at the same time, can be time-sharing batch verification of a variety of thermometers;
◆The temperature field test of verification furnace, oil bath and water bath can be conducted;
◆The temperature of the inexpensive metal thermocouple verification point can choose the temperature point automatically provided by the verification procedure (or recommended), and allows the user to reset the verification point according to the actual use temperature of the couple, the temperature of the verification point can be set arbitrarily, such as: 358.5℃, 772.6℃, etc.;
◆In the calibration of thermocouple, there are three compensation methods: freezing point compensation, manual setting compensation and automatic cold end compensation;
◆High degree of automation of the verification process: the computer from the wire check (judge whether the wiring is correct), temperature field temperature control to the verification process is all automatic;
◆Adopt the standard temperature control principle, hardware and software double temperature control, in the case of computer failure, can choose manual operation, effectively avoid the computer failure and can not continue the verification work.
Main Functions of Software
◆The calibration system software can be operated in Chinese Win10/7/xp and above, etc.;
◆Record forms and certificates such as "verification" or "calibration"; verification certificates or calibration certificates, select standard or custom format; export original data of record forms and certificates to Excel;
◆The software supports a variety of self-diagnosis functions, including temperature field test, scale table query, total uncertainty analysis, line checking and other functions;
◆Support multimedia sound alarm, prompt function, with overtemperature alarm function;
◆With the function of power loss protection and renewal of inspection;
◆The software system displays rich interface, large information, with detailed and vivid real-time curves, data; in the same computer screen can not only observe the temperature control curve of all thermocouple furnace (the curve is clear), but also can switch to a picture to observe the temperature control curve of one thermocouple furnace, convenient and intuitive;
◆The software can complete the functions of automatic verification, automatic data processing and report output, save, query, preview and repeat the verification records, and edit the verification records and verification certificate format.
Temperature Verification System Software
◆In line with the request of the latest national metrological calibration regulations
Accuracy | |||
Six and a Half Digit High-precision Digital Calibrator | Accuracy 0.003%; resolution ratio 0.1μV, 0.1mΩ | ||
Seven and a Half Digit High-precision Digital Calibrator | Accuracy 0.002%; resolution ratio 0.01μV, 0.01mΩ | ||
Eight and a Half Digit High-precision Digital Calibrator | Accuracy 0.0006%; resolution ratio 0.001μV, 0.001mΩ | ||
Scanning Switch | Parasitic potential<0.2 μ V can achieve automatic or manual switching | ||
Compensation Range of Thermocouple Reference Terminal | 5-50℃(resolution ratio 0.1℃) | ||
Temperature Control Stability | |||
Thermocouple Calibration Furnace | 0.1℃/10min | ||
Thermostatic Bath | 0.01℃/10min | ||
Uncertainty | The total uncertainty of thermocouple calibration is better than 0.7 ℃, and the repeatability error is less than 0.25 ℃. The total uncertainty of thermal resistance calibration is better than 50mk, and the repeatability error is less than 10 mk | ||
Item | Model | Specification & Function | Quantity |
1 | YH2000-4 System integration cabinet | Electric measuring meter and multi-channel scanning device, temperature control system integrated cabinet | 1 Unit |
2 | YH2000-1 Multi-channel scanning device | Multi-channel scanning is provided with a parasitic potential of 0.2 μ V | 1 Set |
3 | YH2000-2 intelligent temperature control system | Intelligent temperature control system, temperature field temperature control | 1 Set |
4 | K2010 high-precision electric measuring instrument | Seven and a half digit digital control instrument | 1 Unit |
5 | Lenovo computer | Configuration is not low Inteli7/8G/1T/512M unique display, 19-inch LCD Display, install the legitimate Windows 7 | 1 Unit |
6 | HP1106 Printer | A4 laser | 1 Unit |
7 | YH2000-3 Multiple serial port card | 485/232 conversion card expansion computer serial port | 1 Set |
8 | YHR V1.0 Temperature verification software | Can verify S R、B、K、N、J、E、T;Pt10、Pt100、Cu50、Cu100; Glass liquid thermometer Bimetallic thermometers, pressure thermometers, etc | 1 Set |
9 | WRPB Second-class standard S thermocouple | Calibrator for thermocouple verification | 1 Unit |
10 | WRPB Second-class standard platinum resistance Pt25 | Calibrator for thermal resistence verification | 1 Unit |
11 | YH-10 cryogenic tank | -10~105℃ (antifreeze fluid) | 1 Unit |
12 | YH-300 thermostatic oil bath | 90℃ ~300℃ (silicone oil) | 1 Unit |
13 | YH600L Pipe type calibration furnace (600mm) | Temperature fields from 300℃ to 1200℃ | 1 Unit |
14 | YH-0 0℃ thermostat | Thermocouple cooling end compensates for the temperature field | 1 Unit |
15 | PT100 Grade A cold end resistance | Used for cold end measurement when thermcouple verification | 1 Unit |
16 | YH2000-5-1 thermocouple junction box | Thermocouple dedicated for test connection | 1 Unit |
17 | YH2000-5-2 Thermal resistance junction box | Thermal resistance dedicated for test connection | 1 Unit |
18 | YH600L-F1 Uniform heating block | Used for thermcouple verification | 1 Unit |
19 | YH600L-F2 tubular furnace bracket | Thermocouple verification matching, with the same height as the geometric center line of the thermocouple verification furnace | 1 Set |
20 | Block shaving machine | Electric, ice broken | 1 Unit |
21 | Anti-freeze fluid | Used in conjunction with the cryogenic tank | 25 Litres |
22 | Simethicone | Used in conjunction with the oil tank | 25 Litres |
23 | Compensation lead wire | S, K type | 1 Set |
24 | Quartz glass protective tube | Protective sleeves with different diameter and lengths are configured according to the verification instruments, meet the calibration requirements | 10 pcs |
25 | Binding wire | Used for thermcouple verification | 1 Bundle |
26 | A number of refractory materials | Asbestos, used for the sealing of the calibration furnace mouth | Some |
27 | Test line | Special use for testing (6 meters) | 1 Set |
28 | Refrigerator | 200L, to making ice | 1 Unit |
29 | Exhaust hood | FANGTAI, install above the oil tank | 1 Unit |